Ultisil® XB-C8 -------------Less retentive than XB-C18

The XB-C8 phase is less retentive than XB-C18 phase, useful for strong hydrophobic compounds that are too strongly retained on C18 phase, and for LC/MS applications, where long retention is not desire. When seperating neutral or other highly retained compounds, XB-C8 can save analytical time. However, when seperating polar compounds, XB-C8 column provides differentselectivity than does XB-C18 column.

 

 

Ultisil® XB-C8

Structural Formula  
pH Range 1.5-10.0
Particle Size 3μm, 5μm, 10μm
Surface Area(m²/g) 320(120 Å), 90(300 Å)
Carbon Loading(%) 12(120 Å), 4(300 Å)
USP List L7
Endcapped Yes

 

Comparison of Retention of XB-C18 and XB-C8

Column: 4.6×150mm, 5μm
Mobile Phase: Water/acetonitrile=30/70
Flow rate: 1.0mL/min
Detector: 344nm
Temperature: 25°C
Samples:

1) Uracil

2) Ethylbenzene

3) Propylbenzene

4) Butylbenzene

5) Amylbezene

6) Heptylbenzene

 

Column: 4.6×150mm, 5μm
Mobile Phase: 20mM phosphate(pH 7.0)/methanol=20/80
Flow rate: 1.0mL/min
Detector: 215nm
Temperature: 25°C
Samples:

1) Uracil

2) Ropranolol

3) Ortriptyline

4) Amitryptyline

5) Trimipramine

 

 

Tirucallol and Euphol

Column: Ultisil® XB-C8, 4.6×250mm, 5μm
Mobile Phase: Acetonitrile/water=90/10
Flow rate: 1.0mL/min
Detector: 210nm
Temperature: 30°C
Samples: 10μL

 

Nalbuphine HCl

Column: Ultisil® XB-C8, 4.6×250mm, 5μm
Mobile Phase: A: 0.02mol/L KH2PO4 buffer(pH 6.5)(%)
B: acetonitrile
Gradient Program:

Time(min)

0

15

25

28

30

45

A(%)

80

70

40

40

80

80

B(%)

20

30

60

60

20

20

Flow rate: 1.0mL/min
Detector: 220nm
Temperature: 30°C
Injection Volume: 10μL

 

Analysis of Insulin Detemir

Column: Ultisil® XB-C18, 4.6×250mm, 5μm
Mobile Phase: A: 20g (NH4)2SO4, 900mL water, 100mL
acetonitrile, adjust pH 2.3
B: acetonitrile/water=80/20;
%B=0(0min) , 30(9min), 60(40min)
Flow rate: 1.0mL/min
Detector: 214nm
Temperature: 30°C
Injection Volume: 20μL